DEFIBTECH, L.L.C.; AUTOMATED EXTERNAL DEFIBRILLATORS Back to Search Results
Model Number DBP-1400
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Troubleshooting of the aed with the customer identified that the aed's battery pack was expired with a labeled expiration date of 06/30/2021.The cause of the aed not powering on was related to a lack of maintenance of the aed.As a follow-up with the customer, the proper maintenance of this device was reviewed.
 
Event or Problem Description
A customer reported their aed is flashing red and prompting a "replace battery pack now warning".They reported this did not occur during patient use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS
Manufacturer (Section D)
DEFIBTECH, L.L.C.
741 boston post rd.
suite 201
guilford CT 06437
Manufacturer (Section G)
DEFIBTECH, L.L.C.
14 commercial st.
branford CT 06405
Manufacturer Contact
ian white
741 boston post road
suite 201
guilford, CT 06437
2034536654
MDR Report Key20858612
Report Number3003521780-2024-01362
Device Sequence Number4058488
Product Code MKJ
UDI-Device Identifier00815098020362
UDI-Public(01)00815098020362
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K013896
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date06/30/2021
Device Model NumberDBP-1400
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2024
Initial Report FDA Received Date12/05/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/24/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-