| Model Number |
VTICM5_12.1 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/03/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation - lens work order search: no similar complaint type events reported for units within the same lot.Manufacturer narrative - secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
Il a facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced lens rotation not associated to a low vault.In a separate visit the lens was exchanged for a longer length lens and the problem resolved.It has been noted that the surgeon selected a different lens size than that initially suggested by the icl calculation software.Therefore, there is not enough safety and effectiveness data to support implantation in this patient category.Cause of event was reported as "other".It was also reported that device did not fail to perform as intended.
|
| |
|
Additional Manufacturer Narrative
|
|
H6- device evaluation: lens was returned with reside/debris on product in microcentrifuge vial.Visual inspection found no visible damage to lens.Dimensional inspection found the lens to be within specifications.Claim # (b)(4).
|
| |
|