• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARGON MEDICAL DEVICES, INC. GUIDEWIRE; CARDIAC/PERIPHERAL VASCULAR GUIDEWIRE, SINGLE-USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES, INC. GUIDEWIRE; CARDIAC/PERIPHERAL VASCULAR GUIDEWIRE, SINGLE-USE Back to Search Results
Model Number 395613
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
For g4: pma/510(k): premarket submission number not available/not released as it is exempt from premarket submission.The guidewire involved in the reported complaint will not be returned for investigation because the customer has disposed of it.Therefore, a physical investigation could not be performed, and the root cause could not be determined.
 
Event or Problem Description
During multiple attempts, the customer had difficulty removing the guidewire from the quattro catheter.Three quattro catheters and zoll guidewires were attempted to insert into the femoral vein (right and left sides).Each attempt was difficult and required multiple insertions.The customer reported that the guidewire could not be removed from any of the three catheters, and the catheter had to be removed along with the guidewire each time.Per the customer, the physician's experience with intravascular catheter placement was frequent, and per the physician, there was no increased tortuosity in the vein.No patient injury was reported.The customer saved only one catheter (lot #198059) with guidewire (lot #199681) for investigation.The lot numbers of the two other catheters and the guidewires are unknown, and the customer has disposed of them.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDEWIRE
Common Device Name
CARDIAC/PERIPHERAL VASCULAR GUIDEWIRE, SINGLE-USE
Manufacturer (Section D)
ARGON MEDICAL DEVICES, INC.
7800 dallas pkwy
ste. 200
plano TX 75024
Manufacturer (Section G)
ZOLL CIRCULATION, INC.
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kimthoa sackrison
2000 ringwood ave.
san jose, CA 95131
4084192347
MDR Report Key20859357
Report Number3010617000-2024-00956
Device Sequence Number4058509
Product Code DQX
UDI-Device Identifier00886333211753
UDI-Public00886333211753
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number395613
Device Catalogue Number395613
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/13/2024
Initial Report FDA Received Date12/05/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-