Additional information was provided in d.3., d.9., h.3., h.6.And h.11.The lens was returned positioned correctly in the lens case.A small amount of viscoelastic was observed on the lens.The optic had two scrape marks across the center.This damage was similar in appearance to damage caused by a plunger override.Complaint history and product history records were reviewed, and the documentation indicated the product met release criteria.It is unknown if qualified associated product were used.The reported lens damage was observed.The optic had two scrape marks across the center.This damage was similar in appearance to damage caused by a plunger override.It is unknown if qualified associated product were used.The instructions for use (ifu) instructs: company foldable intraocular lens (iols) are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The handpiece ifu instructs: verify that the plunger tip is properly engaging the lens optic and that the lens moves forward at the same rate as the plunger while slowly advancing the plunger forward to avoid damaging the lens.When the threads on the knob make contact with the barrel, turn the knob clockwise approximately half turn to engage the threads and then stop.The iol will now be in the dwell position.Inspect to ensure the plunger is behind the optic.Due diligence has been performed in an attempt to obtain further information on this event.The facility has not responded to requests for follow-up information.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|