BOSTON SCIENTIFIC CORPORATION FLEXIMA; TUBE, DRAINAGE, SUPRAPUBIC Back to Search Results
Model Number M001274030
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/06/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the cannula and guidewire were difficult to remove.A flexima drainage catheter system and amplatz-pi guidewire were selected for use.During an antegrade ureteric stent stenting, the drainage catheter cannula would not detach and was lodged in the drainage catheter stent.The amplatz-pi guidewire could not be removed easily, and required excessive force to remove it.As a result, the amplatz-pi guidewire frayed and fractured when retracted.The patient was returned to the ward with the stent in place with the cannula attached acting like the drain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIMA
Common Device Name
TUBE, DRAINAGE, SUPRAPUBIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key20864250
Report Number2124215-2024-76493
Device Sequence Number15663130
Product Code FAD
UDI-Device Identifier08714729323785
UDI-Public08714729323785
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K944290
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM001274030
Device Catalogue NumberM001274030
Device Lot Number0033643017
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2024
Initial Report FDA Received Date12/06/2024
Date Device Manufactured03/15/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-