EXACTECH, INC. LOGIC FEMORAL PS CEM LEFT SZ 2; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number LOGIC FEMORAL PS CEM LEFT
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 10/06/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: concomitants: 02-012-35-2013 - logic tibia ps mod insrt sz 2 13mm, (b)(6); 02-012-45-2020 - lgc tibial fit tray cem sz 2f / 2t, (b)(6); 200-02-29 - three peg patella 29mm, (b)(6); 204-34-02 - fluted stem extension 25l x 14 mm, (b)(6); 204-70-00 - tibial stem ext.Screw, (b)(6).The product associated with the reported event is within the scope of recall z-1413-2024; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Event or Problem Description
Legal case-usa: it was reported via legal documentation that a patient had a left total knee arthroplasty on (b)(6) 2017, and then experienced revision surgical procedure on (b)(6) 2023 approximately 6 years and 8 months after initial implant.No images were provided.There is no other information available.No device return anticipated due to this being a legal case.Ebi initial surgery attached.Historical record & smartsolve searched for sn/patient name with no results.Operative report from revision surgery attached.No further information.As directed by qa management: this legal complaint case is from the united states.The event did not occur in, nor is it reportable for brazil, canada, eea/eu, japan, taiwan, or great britain, excluding northern ireland.Therefore, only the us reportability determination will be assessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC FEMORAL PS CEM LEFT SZ 2
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key20864614
Report Number1038671-2024-04663
Device Sequence Number20988120
Product Code JWH
UDI-Device Identifier10885862001122
UDI-Public10885862001122
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/18/2021
Device Model NumberLOGIC FEMORAL PS CEM LEFT
Device Catalogue Number02-010-01-0220
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/16/2023
Initial Report FDA Received Date12/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age70 YR
Patient SexFemale
-
-