STRYKER ENDOSCOPY-SAN JOSE RF 2 PROBES 90-S; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 0279351100
Medical Device Problem Codes Failure to Deliver Energy (1211); Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device had insufficient / no energy delivered.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the device had insufficient / no energy delivered.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: cut function not working probable root cause: probe short (possibly due to fluid ingress), open circuit (possibly due to loose electrical connection), power output variation, wrong default setting, console error crossflow temperature mitigation actions use errors: user attempts to sterilize, disinfect, clean or reuse probe incorrect power level set nonbendable probe bent with probe bender, or a bendable probe bent with anything other than the stryker rf probe bender attempting to use probe without completely immersing tip in a conductive irrigant probe handle is submerged in liquid or suction tube is cut the reported failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RF 2 PROBES 90-S
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key20865155
Report Number0002936485-2024-00920
Device Sequence Number18298289
Product Code GEI
UDI-Device Identifier07613327056952
UDI-Public07613327056952
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K160050
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0279351100
Device Lot Number23269AE2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/07/2024
Supplement Date Received by Manufacturer03/12/2025
Initial Report FDA Received Date12/06/2024
Supplement Report FDA Received Date03/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-