| Model Number |
690R30 |
| Medical Device Problem Code |
Intravalvular regurgitation (4068)
|
| Health Effect - Clinical Code |
Tricuspid Valve Insufficiency/ Regurgitation (4453)
|
| Date of Event |
12/23/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that 5 months post implant of this 30mm tricuspid annuloplasty ring, severe tricuspid regurgitation was noted.It was reported that percutaneous tricuspid valve reconstruction using triclip implantation between the septal and posterior tricuspid native valve leaflet was performed.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received additional information which stated that approximately 3 weeks prior to the percutaneous tricuspid valve reconstr uction being performed, severe tricuspid regurgitation had been identified on a echocardiogram.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|