| Model Number |
MA60AC |
| Medical Device Problem Codes |
Defective Device (2588); Failure to Eject (4010)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported with the description of lens stuck in cartridge, would not advance, bonded lens.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Fda product problem code a0203 removed the lens was returned posterior surface up in a lens case.Viscoelastic was dried on the lens.The lens was cleaned with klrp.No lens damage was observed.The lens dimensions were acceptable using an approved template.The cartridge was not returned.Complaint and product history records were reviewed, and documentation indicated the product met release criteria.Associated products were not provided.It is unknown if qualified associated products were used.The company lens is only qualified for use with the company ii a and ii b cartridges.The root cause for the reported lens stuck in cartridge could not be determined.The lens was returned in the lens case.No damage was observed.The lens dimensions were acceptable using an approved template.The cartridge was not returned.It is unknown if a qualified cartridge was used.The company lens is only qualified for use with the company ii a and ii b cartridges.The instructions for use (ifu) instructs: company foldable intraocular lenses (iols) are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The ifu instructs: use holding forceps to grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until it is centered with or slightly past the outline etched into the top of the cartridge.The trailing haptic will extend from the proximal end of the cartridge.Position the trailing haptic to the left of the haptic post as shown in the detailed view.This will allow the plunger to go past the haptic during the initial advancement of the plunger into the cartridge.Verify the lens is positioned on the bottom surface of the cartridge.The haptic should be maintained to the left of the post when the lens is loaded correctly.Accurate positioning of the lens will decrease the potential for optic and haptic damage.Due diligence has been performed in an attempt to obtain further information on this event.The facility has not responded to requests for follow-up information.The file will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|