CORDIS US. CORP SABER; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Catalog Number 48003004S
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
As reported, the balloon of 3mm x 4cm saber percutaneous transluminal angioplasty (pta) balloon catheter burst during the procedure.There were no reported injuries to the patient.The device was stored and prepped per the instructions for use (ifu) and there were no visible signs of damage to the device or device packaging.There was no difficulty removing any of the sterile packaging components.The device maintained negative pressure during preparation.The physician achieved certification on the use of the saber device prior to the procedure.The target vessel was the superficial femoral artery (sfa).The target site vessel characteristics included 80% stenosis, with no calcification and no tortuosity.The device will be returned for evaluation.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections have been updated accordingly: b4, d9, g3, g6, h1, h2, h3, and h11 this device is available for analysis, but the engineering report is not yet available.However, it will be submitted within 30 days upon receipt.Additional information is pending and will be submitted within 30 days upon receipt.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections have been updated accordingly: b4, g3, g6, h1, h2, h3, h6, and h11 complaint conclusion: as reported, the balloon of 3mm x 4cm saber percutaneous transluminal angioplasty (pta) balloon catheter burst during the procedure.There were no reported injuries to the patient.The device was stored and prepped per the instructions for use (ifu) and there were no visible signs of damage to the device or device packaging.There was no difficulty removing any of the sterile packaging components.The device maintained negative pressure during preparation.The physician achieved certification on the use of the saber device prior to the procedure.The target vessel was the superficial femoral artery (sfa).The target site vessel characteristics included 80% stenosis, with no calcification and no tortuosity.A non-sterile unit of ¿saber 3mm4cm 90¿ was received for analysis coiled inside of a clear plastic bag.The device was unpacked to proceed with the product evaluation.A thorough inspection was performed observing that the unit does not present any damages or anomalies.The balloon arrived not inflated.No other outstanding details were observed.Functional test was performed.One inflator/deflator device was attached to the inflation port.Balloon inflation test was done applying positive pressure using the inflator/deflator device with the purpose of reaching the rate burst pressure.However, inflation pressure was not maintained due to a leak in the balloon caused by a rupture on the surface.The balloon was inspected with a vision system observing a rupture on the surface where the water was leaking.The balloon surface presented evidence of a ruptured condition and secondary scratch marks located near the damaged border area.The failure reported by the customer as ¿balloon~burst¿ was confirmed.A ruptured condition was observed on the balloon surface that generated a leak, which impeded maintenance of the inflation pressure.The exact cause of this condition observed on the balloon could not be conclusive determined during the analysis.Procedural factors and handling process may have contributed to the failure as reported, such as the stenosis reported.This type of damage is commonly caused during the interaction of the material with a sharp object or mechanical damage.It is very likely that the same factors that caused the observed scratch marks on the surface could probably led to the damaged condition found on the received device.It seems that the material near the damaged area was affected with a sharp object from the outside of the device.The product analysis does not suggest that the observed damage could be related to the manufacturing process; therefore, no corrective action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SABER
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS US. CORP
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS US CORP.
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
coyol free zone, building b25, ,  , costa rica
el coyol alajuela bld 25 
7863138372
MDR Report Key20865861
Report Number3007635982-2024-00282
Device Sequence Number8491184
Product Code LIT
UDI-Device Identifier20705032068069
UDI-Public(01)20705032068069(17)250731(10)82255895
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K971010
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date07/31/2025
Device Catalogue Number48003004S
Device Lot Number82255895
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2024
Supplement Date Received by Manufacturer12/06/2024
01/13/2025
Initial Report FDA Received Date12/06/2024
Supplement Report FDA Received Date12/19/2024
01/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.
Patient SexUnknown
-
-