| Model Number |
1012270-20 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.Based on the information received, the investigation determined that the reported complaints appear to be related to operational context.There was no leak noted during preparation, which suggests a product quality issue did not contribute to the reported difficulties.It was reported by the account that the balloon dilatation catheter (bdc) interacted with the mildly calcified and moderately tortuous anatomy resulting in the reported difficulty advancing the device and subsequently the balloon became damaged and ruptured during inflation.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a 90% stenosed de novo lesion in the distal right coronary artery (drca) with mild calcification and moderate tortuosity.The 2x20mm mini trek balloon dilatation catheter (bdc) was advanced to the target lesion with resistance due to the anatomy.The balloon was inflated; however, a ruptured occurred during the first inflation at 6 atmospheres (atm).Therefore, the bdc was removed from the patient.Another trek balloon was used to continue the procedure.There are no adverse patient effects and no clinically significant delays reported in the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|