• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. 2008T HD SYS. CDX BLUE STAR; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. 2008T HD SYS. CDX BLUE STAR; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 191124
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 12/02/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
A user facility's biomedical technician (bmt) reported that a 2008t hemodialysis (hd) machine's blood pump rotor roller guide pin cut a bloodline 1 hour into a patient's hd treatment.The machine, a fresenius 2008t machine, alarmed appropriately with an "air in line" alarm.A fresenius dialyzer was also in use.The patient¿s estimated blood loss (ebl) was approximately 500 ml.There was no patient injury or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.The bloodline was discarded.Follow up with the bmt revealed that that the machine has not been returned to service pending a replacement blood pump rotor.No sample is available.
 
Event or Problem Description
A user facility's biomedical technician (bmt) reported that a 2008t hemodialysis (hd) machine's blood pump rotor roller guide pin cut a bloodline 1 hour into a patient's hd treatment.The machine, a fresenius 2008t machine, alarmed appropriately with an ¿air in line¿ alarm.A fresenius dialyzer was also in use.The patient¿s estimated blood loss (ebl) was approximately 500ml.There was no patient injury or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.The bloodline was discarded.Follow up with the bmt revealed that that the machine has not been returned to service pending a replacement blood pump rotor.No sample is available.
 
Additional Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HD SYS. CDX BLUE STAR
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key20868101
Report Number0008030665-2024-01200
Device Sequence Number18283860
Product Code KDI
UDI-Device Identifier00840861102082
UDI-Public00840861102082
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K173972
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 05/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number191124
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2024
Supplement Date Received by Manufacturer05/06/2025
Initial Report FDA Received Date12/06/2024
Supplement Report FDA Received Date05/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/13/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRESENIUS BLOODLINES; FRESENIUS BLOODLINES; FRESENIUS DIALYZER; FRESENIUS DIALYZER
Patient Age54 YR
Patient SexMale
Patient Weight113 KG
-
-