|
On 11/06/2024, it was reported by an arthrex subsidiary employee via (b)(4) that an ar-6410 main pump tubing mbrane was crushed, causing the pressure to become abnormally high and the patient's leg became swollen.This was discovered during use in a case.The surgeon stopped using the product and the irrigation was performed by (b)(4).Additional information requested to the rep.
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to misuse/mishandling due to damage to the device.
|