| Model Number |
6123 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a captivator ii-15 mm round stiff snare was used in the colon during a polypectomy procedure for colorectal polyps performed on (b)(6) 2024.During the procedure, it was reported that the metal fittings on the active cord and the gold metal fittings had come off.The procedure was completed with another captivator ii-15 mm round stiff snare device.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0501 captures the reportable event of the cautery pin detached.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a captivator ii-15mm round stiff snare was used in the colon during a polypectomy procedure for colorectal polyps performed on (b)(6) 2024.During the procedure, it was reported that the metal fittings on the active cord and the gold metal fittings had come off.The procedure was completed with another captivator ii-15mm round stiff snare device.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0501 captures the reportable event of the cautery pin detached.Block h11: (product investigation) the returned captivator snare was analyzed, and visual and microscope inspection that the 2-in-1 connector was broken and detached.No other device problems were noted.The reported complaint of cautery pin detachment of device or device component was confirmed since the 2-1 pin was broken and detached.Based on the information available and the device analysis performed, the most probable root cause for the reported complaint is manufacturing deficiency.Boston scientific has initiated an investigation to further evaluate cautery pin detachment events to determine root cause of these events, as well as appropriate mitigation(s).
|
| |
|