MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES¿ 90K; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES¿ 90K; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-02H
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 11/25/2024
Type of Reportable Event Malfunction
Event or Problem Description
The recipient reportedly experienced sound quality issues.Device testing revealed results within normal limits.The external device was noted as working.Programming adjustments were made; however, the issue did not resolve.The recipient's device was explanted.The recipient was reimplanted with another cochlear device.
 
Additional Manufacturer Narrative
A review of the device history record was completed and no anomalies were noted.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Additional Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The recipient's device will not return to advanced bionics for analysis.A review of the device history record was completed and no anomalies were noted.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ 90K
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
laurel masamitsu
28515 westinghouse place
valencia, CA 91355
MDR Report Key20872428
Report Number3006556115-2024-01865
Device Sequence Number15663115
Product Code MCM
UDI-Device Identifier07630016816260
UDI-Public(01)07630016816260(11)150713(17)170630
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P960058
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/30/2017
Device Model NumberCI-1400-02H
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/06/2024
Supplement Report FDA Received Date01/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/13/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age11 YR
Patient SexMale
-
-