SMITH & NEPHEW, INC. ULTRA FAST-FIX AB ASSEMBLY - CURVED; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
|
Back to Search Results |
|
| Catalog Number |
72201494 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during an arthroscopy, the t2 of the ultra fast-fix did not trigger.Surgery was completed with a smith and nephew back-up device instead.There was a delay of less than 30 minutes, and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h2: additional information in d4(lot number and expiration date) and h4.
|
| |
|
Additional Manufacturer Narrative
|
|
H10.H3, h6: the reported device was received for evaluation.A visual inspection revealed the device was returned in original packaging with the batch number of the complaint on the label.The t1 is off the device but attached to the suture.The t2 and suture were returned on the device.The variable depth straw was not returned.Bio debris is present.A functional evaluation was performed on the returned device and found that the actuator does not extend far enough to fully deploy the t2 off the needle.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences.Other factors, that can contribute to the complaint event, include an application of unintended inappropriate or excessive force to the device.Based on this investigation, there is no evidence to suggest a design, material or manufacturing issue.Therefore, the need for corrective action is not indicated.
|
| |
|