| Model Number |
MI1210 SYNCHRONY ST |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected.One week before, the user removed his audio processor as it was making strange noises that he could not describe.No trauma was reported and no redness or swelling over the implant area were noticed.Re-implantation is being considered.No date has been scheduled yet.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress seems very likely.However to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected.Re-implantation is being considered.No date has been scheduled yet.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress seems very likely.However, to determine an exact root cause device investigation would be necessary.The concerned device was explanted but has not been received for investigation yet.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected.The user was re-implanted on (b)(6) 2025.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected.The user was re-implanted on 11-jun-2025.
|
| |
|