ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065753135
Medical Device Problem Code Suction Failure (4039)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/30/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that the aspiration failure occurred during ultra sound.The surgery was completed on the same day after replacing the product with another one.There was no impact to the patient.
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The turbid active fluidics management system in the tray was visually inspected.The drain bag was detached from the drain bag tape on the cover due to saturation.The drain bag tape was adhered on the cassette properly.Both irrigation and aspiration luers were intact.Flare shape was observed in the blue socket fitting at the aspiration tubing insertion end.The sample was tested using a calibrated console with the current software revision.The sample could be primed and tuned, and the service data could be retrieved from the console.The aspiration and irrigation pressures met specification.The root cause of the customer's complaint could not be established since the returned sample met specifications.No contributing factors could be identified that could cause the reported complaint.After the investigation of this complaint, it has been determined that this sample met specifications.Therefore, no action will be taken at this time.Based on our current tracking, there are no adverse trends for this reported complaint.Quality assurance will continue to monitor customer complaints via the complaint review meetings and will take action for any future occurrences as is deemed necessary.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20876151
Report Number1644019-2024-02954
Device Sequence Number18063111
Product Code LRO
UDI-Device Identifier00380657531356
UDI-Public(01)00380657531356(17)260630(11)240731(10)170A6C
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Catalogue Number8065753135
Device Lot Number170A6C
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/14/2024
Supplement Date Received by Manufacturer02/13/2025
Initial Report FDA Received Date12/09/2024
Supplement Report FDA Received Date03/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/31/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CENTURION VISION SYSTEM (ACTIVE SENTRY)
Patient SexUnknown
-
-