MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® TEMPERATURE MANAGEMENT SYSTEM, ARCTICGEL¿ PADS; ARCTIC GEL PADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® TEMPERATURE MANAGEMENT SYSTEM, ARCTICGEL¿ PADS; ARCTIC GEL PADS Back to Search Results
Catalog Number 318-01
Medical Device Problem Codes Gas/Air Leak (2946); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/06/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.The complete initial reporter phone number is (b)(6).H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that pad leak velocity is low.
 
Additional Manufacturer Narrative
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The labeling is found to be adequate, and it states, ¿select the proper number of pads for the patient size and clinical indication.However, the rate of temperature change and potentially the final achievable temperature is affected by pad surface area, patient size, pad placement and water temperature range.Best system performance will be achieved by using the maximum number of pads.¿place the pads on healthy, clean skin only.Locate pad edges away from articulating areas of the body to avoid irritation.Place pads to allow for full respiratory excursion.¿the pad surface must be contacting the skin for optimal energy transfer efficiency.The small universal pads are provided with a cloth liner over the hydrogel.The pads may be used with the cloth liner in place or removed to expose the hydrogel adhesive.If the cloth liner is used, secure the pads with the velcro straps to ensure good contact with the skin.The pads may be removed and reapplied if necessary.¿attach the pad¿s line connectors to the fluid delivery line manifolds.Begin treating the patient.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the arctic gel pad leak velocity was low.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC SUN® TEMPERATURE MANAGEMENT SYSTEM, ARCTICGEL¿ PADS
Common Device Name
ARCTIC GEL PADS
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key20877862
Report Number1018233-2024-07834
Device Sequence Number18093989
Product Code DWJ
UDI-Device Identifier00801741080128
UDI-Public(01)00801741080128(11)220330(17)240331(10)NGGQ1698
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K142702
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Catalogue Number318-01
Device Lot NumberNGGQ1698
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2024
Supplement Date Received by Manufacturer02/10/2025
Initial Report FDA Received Date12/09/2024
Supplement Report FDA Received Date02/11/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-