Model Number 0276
Medical Device Problem Code Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the health care professional (hcp) called technical services (ts) for further review of this cardiac resynchronization therapy defibrillator (crt-d) stored ventricular episodes due to concerns of oversensing on the right ventricular (rv) lead channel.Upon review of the stored ventricular episode, ts observed that on the right atrial (ra) channel, the patient appeared to be in an atrial flutter (af) arrhythmia.Further review of the episodes, ts determined that there was crosstalk oversensing of the af on the rv lead.The hcp noted that the patient is on antiarrhythmic medication to treat the patient underlying arrhythmia and has been in normal sinus since the ventricular episode.Ts discussed reprogramming options with the hcp.At this time, the crt-d and rv lead remain in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key20879232
Report Number2124215-2024-77571
Device Sequence Number12940755
Product Code NVY
UDI-Device Identifier00802526531231
UDI-Public00802526531231
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P910073
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/12/2021
Device Model Number0276
Device Catalogue Number0276
Device Lot Number420802
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2024
Initial Report FDA Received Date12/09/2024
Date Device Manufactured03/12/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexMale
-
-