• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION GEM MICROVASCULAR ANASTOMOTIC COUPLER; DEVICE,ANASTOMOTIC,MICROVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION GEM MICROVASCULAR ANASTOMOTIC COUPLER; DEVICE,ANASTOMOTIC,MICROVASCULAR Back to Search Results
Model Number GEM2753
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/07/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the microvascular anastomotic coupler device ring ¿fall itself¿.This issue occurred pre-operation.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
E1: initial reporter address - (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Additional information: d9, h3, h6, h11.H11: the actual device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.There appeared to be no issues with the returned coupler rings, however, it is unknown what may have transpired during preparation for the surgical process that may have caused the ring to become dislodged.There is potential that accidental undue pressure placed on the ring during preparation for use may have caused the ring to dislodge from the jaw assembly.The reported condition could not be confirmed or refuted, and the cause of the reported condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GEM MICROVASCULAR ANASTOMOTIC COUPLER
Common Device Name
DEVICE,ANASTOMOTIC,MICROVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
SYNOVIS SURGICAL INNOVATIONS
2575 university ave w
saint paul MN 55114
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key20879752
Report Number1416980-2024-07104
Device Sequence Number8972313
Product Code MVR
UDI-Device Identifier00844735002669
UDI-Public(01)00844735002669(10)SP24C072034089
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K861985
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberGEM2753
Device Catalogue Number511100250060
Device Lot NumberSP24C072034089
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2024
Supplement Date Received by Manufacturer12/20/2024
Initial Report FDA Received Date12/09/2024
Supplement Report FDA Received Date01/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-