DEPUY MITEK LLC US VERSALOK THREADER TAB *EA; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 210800
Medical Device Problem Codes Device-Device Incompatibility (2919); Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device was discarded by the customer, therefore not returned and unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number and no non-conformance was identified.Based on the current available information, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Event or Problem Description
It was reported that during an unspecified surgical procedure for shoulder, it was found that the five (x5) versalok threader tab devices were being used when the surgeon attempted to deploy the anchor in question with a deployment gun, but the anchor was not stable and came loose after the fire.Four anchors were then used, but a similar event occurred and all five anchors failed to deploy in the body and came loose.With another implant, the tear was sutured and the surgery was completed.There was 60 minutes surgical delay and no harm to the patient.The exact date of the event was unknown.However, it was reported that the event occurred in 2024.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.This is report 4 of 5 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key20881008
Report Number1221934-2024-04374
Device Sequence Number21025986
Product Code HWC
UDI-Device Identifier10886705001163
UDI-Public(01)10886705001163(10)169L228(17)260531
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K063478
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Catalogue Number210800
Device Lot Number169L228
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2024
Initial Report FDA Received Date12/09/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/08/2023
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-