EXACTECH, INC. LOGIC FEMORAL PS CEM RIGHT SZ 2; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
9999 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Unstable (1667); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Synovitis (2094); Osteolysis (2377); Joint Laxity (4526); Implant Pain (4561); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
06/06/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Report number: 1038671-2024-00235: d10 concomitants: (b)(6); 02-012-35-2015 - logic tibia ps mod insrt sz 2 15mm.(b)(6); 02-012-45-2020 - lgc tibial fit tray cem sz 2f / 2t.(b)(6); 200-03-35 - one peg patella 35mm.Multiple mdr reports were filed for this event; associated reports: 1038671-2024-00235.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, and loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported via legal documentation that a patient had a right total knee arthroplasty on (b)(6) 2016, and then experienced a revision surgical procedure on (b)(6) 2022, approximately 6 years, 2 months after initial implant.Patient experienced prosthesis wear, loosening, osteolysis, synovitis, joint laxity, failure of implant and implant pain.No further issues or complications were reported.There was no other patient/medical information provided.No x-rays or images were provided.The device will not be returned.There is no other information available.
|
| |
|
Additional Manufacturer Narrative
|
|
The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, and loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.D4: corrected.H6: corrected health effect and medical device clinical codes.
|
| |
|