Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use of the bd vacutainer® sodium fluoride: 3 mg na2 edta: 6 mg blood collection tubes, twenty (20) tubes were observed to be missing additive.There were no health impact or consequences reported.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use of the bd vacutainer® sodium fluoride: 3 mg na2 edta: 6 mg blood collection tubes, twenty (20) tubes were observed to be missing additive.There were no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d9: device available for evaluation: yes.H.3 device eval by manufacturer? yes.D9: returned to manufacturer on: 12-dec-2024.Investigation summary: bd received twenty (20) samples and one (1) photo for investigation.The photo was reviewed and the indicated failure mode for missing additive was observed.Additionally, the customer samples were evaluated by visual examination and the issue of missing additive was observed.Retention samples were evaluated by visual examination and the issue of missing additive was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode of missing additive.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|