| Model Number |
407694 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Related manufacturing ref: 3005334138-2024-00526, 3005334138-2024-00527, 3005334138-2024-00528, 3005334138-2024-00530.During the procedure, the introducer was inserted and the guide catheter could not be passed over the valve.Four additional introducers from the same lot were used which had the same issue.A non abbott introducer was used to complete the procedure with no consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h2, h3, h6.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported insertion difficulties and subsequent delay remains unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
One 6f ultimum hemostasis introducer sheath and dilator were received for evaluation.The guidewire assembly was also returned.The reported event of guidewire insertion difficulty could not be confirmed.No resistance or other functional anomalies were noted during guidewire insertion through the dilator.Both the dilator distal tip inside diameter and guidewire outside diameter measurements were within specifications.The batch record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported insertion difficulty remains unknown.
|
| |
|