ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA MEDIAL CONGRUENT ARTICULAR SURFACE RIGHT 11MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Damage to Ligament(s) (1952); Pain (1994); Scar Tissue (2060); Swelling/ Edema (4577); Joint Impingement (4834)
Date of Event 02/19/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10 - concomitant devices - persona cemented cruciate retaining standard femoral component right size 10 catalog#: 42502606802, lot#: 65328773, persona cemented stemmed tibial component right size f catalog#: 42532007502, lot#: 65440949, persona cemented all poly patella 32mm catalog#: 42540000032, lot #: 65737320.The complainant has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient underwent a right knee arthroplasty revision approximately one (1) year post-operatively to address pain, instability, swelling, quadriceps contraction and posterior cruciate ligament (pcl) deficiency with impingement.Initial operative notes noted no intraoperative complications.Revision operative notes noted the patient was diagnosed with global instability with pcl deficiency and severe arthrofibrosis with impingement and hypertrophic scar.The patient had persistent pain, instability and swelling since initial surgery with quadriceps contraction and unable to perform straight leg raise without lag.There was a deficiency of the posterior cruciate ligament with flexion instability.Inspection revealed the implants were well-positioned and well-fixed with no signs of poly wear or loosening.There was significant scarring within the joint space, suprapatellar space, gutters, as well as encapsulation of almost the entire polyethylene of the patella.All components were removed and replaced except the patella without complications.Attempts have been made; however, no additional information is available.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: b4, g3, g6, h2, h3, h6, h11.The following sections were corrected: h4.Medical records provided confirmed the reported event.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA MEDIAL CONGRUENT ARTICULAR SURFACE RIGHT 11MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key20884056
Report Number3007963827-2024-00491
Device Sequence Number8934151
Product Code MBH
UDI-Device Identifier00889024468429
UDI-Public(01)00889024468429(17)271108(10)65552664
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K150090
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42522100811
Device Lot Number65552664
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2024
Supplement Date Received by Manufacturer04/30/2025
Initial Report FDA Received Date12/09/2024
Supplement Report FDA Received Date04/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-