• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALERE SAN DIEGO, INC. SURE-VUE URINE HCG; VISUAL, PREGNANCY HCG, PRESCRIPTION USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SAN DIEGO, INC. SURE-VUE URINE HCG; VISUAL, PREGNANCY HCG, PRESCRIPTION USE Back to Search Results
Model Number FHC-102
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3 - date of event is an estimate as the actual date of occurrence was not provided.Preliminary investigation conclusion (pending completion of return testing): retained devices from the reported lot number were tested with hcg-negative clinical urine samples and low-hcg urine standards (3 miu/ml).The results were read at 3 and 4 minutes and all devices yielded the expected negative results.No false positive results were observed during in-house testing.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Manufacturing batch record review did not uncover any relevant non-conformances and found that the lot met quality control specifications.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retained product.Complaints are tracked and trended on a monthly basis.Per the package insert: very low levels of hcg (less than 50 miu/ml) are present in urine specimen shortly after implantation.However, because a significant number of first trimester pregnancies terminate for natural reasons, a test result that is weakly positive should be confirmed by retesting with a first morning urine specimen collected 48 hours later.A number of conditions other than pregnancy, including trophoblastic disease and certain non-trophoblastic neoplasms including testicular tumors, prostate cancer, breast cancer, and lung cancer, cause elevated levels of hcg.Therefore, the presence of hcg in urine specimen should not be used to diagnose pregnancy unless these conditions have been ruled out.This test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.
 
Event or Problem Description
The customer reported a false positive result when testing a patient sample with the sure-vue urine hcg test.The customer was unable to provide any additional information.There was no adverse impact to the patient as a result of the false positive results.The customer also reported false positive results for two other patients.See h10 for the related mdr numbers.
 
Event or Problem Description
The customer reported a false positive result when testing a patient sample with the sure-vue urine hcg test.The customer was unable to provide any additional information.There was no adverse impact to the patient as a result of the false positive results.The customer also reported false positive results for two other patients.See h10 for the related mdr numbers.
 
Additional Manufacturer Narrative
B3 - date of event is an estimate as the actual date of occurrence was not provided.D9, h3, h6, and investigation conclusion (below) updated to reflect receipt and testing of returned devices.Investigation conclusion: retained and returned devices from the reported lot number were tested with hcg-negative clinical urine samples and low-hcg (3 miu/ml) negative urine standards.The results were read at 3 and 4 minutes and all devices yielded the expected negative results.No false positive results were observed during in-house testing.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Manufacturing batch record review did not uncover any relevant non-conformances and found that the lot met quality control specifications.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.A root cause could not be determined from the available information as the reported issue was not replicated during testing of either retained or returned product.Complaints are tracked and trended on a monthly basis.Per the package insert: very low levels of hcg (less than 50 miu/ml) are present in urine specimen shortly after implantation.However, because a significant number of first trimester pregnancies terminate for natural reasons, a test result that is weakly positive should be confirmed by retesting with a first morning urine specimen collected 48 hours later.A number of conditions other than pregnancy, including trophoblastic disease and certain non-trophoblastic neoplasms including testicular tumors, prostate cancer, breast cancer, and lung cancer, cause elevated levels of hcg.Therefore, the presence of hcg in urine specimen should not be used to diagnose pregnancy unless these conditions have been ruled out.This test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURE-VUE URINE HCG
Common Device Name
VISUAL, PREGNANCY HCG, PRESCRIPTION USE
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9942 mesa rim rd
san diego CA 92121
Manufacturer Contact
monika burrell
9942 mesa rim rd
san diego, CA 92121
8588052506
MDR Report Key20886173
Report Number2027969-2024-00215
Device Sequence Number8710064
Product Code JHI
UDI-Device Identifier00613647000526
UDI-Public(01)00613647000526(17)260114(10)0000833012
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K993317
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/14/2026
Device Model NumberFHC-102
Device Catalogue Number23900526
Device Lot Number0000833012
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2024
Supplement Date Received by Manufacturer03/13/2025
Initial Report FDA Received Date12/09/2024
Supplement Report FDA Received Date03/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age27 YR
Patient SexFemale
-
-