| Model Number |
2342306M |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis # (b)(4): part # 2342306m; lot # ct16l076 visual and optical examination confirmed the instrument tip has broken and the corners of the torx tip have been rounded off.The remaining tip has also been twisted.This type of damage is consistent with torsional overload.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from user facility via manufacturing representative regarding a product used in transforaminal lumbar inter body fusion (tlif).It was reported that tip of driver is bent and not holding screw straight.No patinet was involved in this event.No further complicatgions were reported/anticipated.
|
| |
|
Search Alerts/Recalls
|