| Model Number |
LS96360-023 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use of a bio-medicus life support cannula, upon opening the product the customer experienced difficulty with the introducer, which could not be removed from the catheter body due to it being sticky/high friction at a specific point. the cannula body was wetted with saline prior to introducing the white introducer, following customer protocol. the introducer could only be retracted by approximately 7-10 cm before sticking. the device was replaced with another manufacturer's product. there was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
B.5.Medtronic received additional information that upon visual inspection of the cannula, the customer noted that when the introducer was removed from the cannula, the cannula tip was not a perfect circle.The external packaging and the protective tube holding the cannula for shipping did not show damage.The customer decided to check other multistage femoral venous cannula in their stock.They observed the same ¿non-perfect circle¿ on the cannula body tip on most of the cannulae in their stock.They opened two expired devices and checked if the introducer was sticky upon removing around 7-10cm.It was confirmed they were sticky.This was noted on the extracorporeal membrane oxygenation (ecmo) life support range as well as the bio-medicus next gen multistage cannulae.They also removed from the hospital an expired multi and also an expired old gen bi-caval and they were both ¿slightly sticky¿ at the 7-10cm area when removing the introducer.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic received additional information that upon visual inspection of the cannula, the customer noted that when the introducer was removed from the cannula, the cannula tip was not a perfect circle.The external packaging and the protective tube holding the cannula for shipping did not show damage.Medtronic received additional information that no tips were obstructed.Once the introducer was inserted, the cannula appeared normal.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: visual inspection showed no outward signs of any damage.The introducer outer diameter (od) was measured using a keyence scope and the cannula tip inner diameter (id) was measured using a plug gage.The introducer od was measured, and it was within specification.The cannula tip id was measured, and it was within specification.The introducer was inserted and removed several times with no issues noted.The reason for return was not confirmed.Correction d3 (mfr name) d3.6 (postal code): these fields have been updated.Correction additional codes (imf (annex f) health impact): this code has been updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|