Product analysis: at analysis it was determined that the programmer power supply electrical resistance tested out of specification.It was noted that the media bay door was broken, the fan required replacement, cosmetic damage was observed on the stylus, the handle cover and the overlay bezel were cracked.It was further noted that the power cord bay and the printer tray were damaged.All found defective parts were replaced and all other identified issues were resolved.Reconfigured hard drive, reloaded and updated software and the programmer then passed all final functional and systems tests.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|