Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/18/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D:10 - 161470, oxf twin-peg cmntd fem lg pma, lot 084400 or 747600.154726, oxf uni tib tray sz e lm pma, lot 938340.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that the patient underwent revision surgery on the left knee seven years and three months post implantation due to unknown cause.All components were exchanged with addition of patella resurfacing.No operative records were provided.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: b4, b5, d10, g3, g6, h2, h3, h6, h11.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations.Review of the complaint histories identified additional similar complaints for the reported items and no additional similar complaints for the reported part and lot combinations.Radiographs were provided and reviewed by a radiologist.The review of the ap view of the bilateral knees demonstrates bilateral medial compartment hemiarthroplasties.The component on the right is appropriately positioned.There is an oblique appearance of the femoral portion of the left knee hemi arthroplasty, which could indicate malposition.There is associated narrowing of the medial joint space of the left knee, which could indicate underlying polyethylene wear.Based on the available information, a definitive root cause could not be determined.However, the radiograph review does mention possible femoral malposition and bearing wear; however, this cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXFORD ANAT BRG LT LG SIZE 3 PMA
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, UNICOMPARTMENTA
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key20892427
Report Number3002806535-2024-00459
Device Sequence Number15686876
Product Code NRA
UDI-Device Identifier05019279786008
UDI-Public(01)05019279786008(17)220724(10)313160
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P010014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date07/24/2022
Device Model NumberN/A
Device Catalogue Number159554
Device Lot Number313160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2024
Supplement Date Received by Manufacturer12/20/2024
Initial Report FDA Received Date12/10/2024
Supplement Report FDA Received Date12/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/24/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE; UNKNOWN PALACOS BONE CEMENT
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
-
-