DEPUY SYNTHES PRODUCTS LLC 11CM ANGLE ATTACHMENT; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC)
|
Back to Search Results |
|
| Catalog Number |
QD11 |
| Medical Device Problem Codes |
Noise, Audible (3273); Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The actual device has been returned and is currently pending evaluation.Once the investigation has been completed, and if additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
Report 2 of 3 for (b)(4).It was reported that (x3) angle attachment devices generated heat and produced abnormal noise.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2024.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Correction: device availability: d9: the device availability date was incorrect in the initial report.The date has been updated from 12/5/2024 to 12/9/2024.Device evaluation: the actual device was returned for evaluation.During repair, it was determined that the reported condition that the device generated heat and produced abnormal noise was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was determined that the device could not insert cutter, was loose and had component damage.It was further determined that the device failed pretest for visual assessment, thimble set screw and cutter insertion.It was noted that the device had shaft flaw, set screw loose and bar insertion was not possible.The assignable root cause was determined to be due to component failure from wear.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|