• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US QUICKSET TAPER HEX SCDR RIGID; HIP INSTRUMENTS : SCREWDRIVERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US QUICKSET TAPER HEX SCDR RIGID; HIP INSTRUMENTS : SCREWDRIVERS Back to Search Results
Catalog Number 227447000
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Deformation (2976)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the straight screw inserter shaft was worn at the tip and would not thread the trial liner.
 
Additional Manufacturer Narrative
Depuy orthopaedics is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy orthopaedics has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy orthopaedics or its employees that the report constitutes an admission that the device, depuy orthopaedics, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h3, h6: the product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned device revealed that the quickset taper hex scdr rigid tip was cross-threaded.No other damage was observed on the device.The device exhibits damage consistent with repeated use.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A functional test was not performed since the mating device was not returned, however, based on the observed condition of the device, it is reasonable to conclude that this condition would likely impede the device from securely attach or holding its mating device.Therefore, the reported allegation can be confirmed.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the quickset taper hex scdr rigid would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET TAPER HEX SCDR RIGID
Common Device Name
HIP INSTRUMENTS : SCREWDRIVERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key20893154
Report Number1818910-2024-25133
Device Sequence Number21025963
Product Code HXX
UDI-Device Identifier10603295109235
UDI-Public(01)10603295109235(10)AG0411
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number227447000
Device Lot NumberAG0411
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/06/2024
Initial Report FDA Received Date12/10/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/15/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-