• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. 145-DEG PE 36MM HUM LINER +0; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. 145-DEG PE 36MM HUM LINER +0; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 320-36-00
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Abscess (1690); Unspecified Infection (1930)
Date of Event 11/18/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Experience report usa.It was reported that this patient's right knee was revised.Since the index surgery, the patient had a colostomy placed.At some point after, patient developed an infection.It is unclear if her infection is related to the colostomy procedure.Patient had an abscess within the soft tissue of her deltoid.Surgeon drained the abscess and performed a one-stage revision.He explanted all the original hardware and implanted a new small superior augment baseplate, blue, red, and black screws and caps, 40 expanded sphere, locking screw, cemented 12mm stem, 0 tray, reverse torque screw and a 2.5mm liner.Original liner had some wear already.Patient was last known to be in stable condition following the event.Is the reported event associated with revision of exactech implants? yes.What side was effected? right.Do you know the initial implant date (index surgery date)? known.Provide the initial implant date (index surgery date) (b)(6)2023.Was the patient revised to exactech devices? yes.Please describe below: dr.(b)(6) performed a one-stage revision.He explanted all the original hardware and implanted a new small superior augment baseplate, blue, red, and black screws and caps, 40 expanded spheres, locking screw, cemented 12mm stem, 0 tray, reverse torque screw and a 2.5mm liner.Original liner had some wear already.Is the reported event related to the breakage of a device? no.Did the reported event lead to a surgical delay/prolongation? yes.How long was the duration (in minutes): > 45 mins did the patient experience any adverse events as a result of the surgical delay/prolongation? no.Please describe below: delay in the sense of the need for revision surgery at all.Patient was last known to be in stable condition following the event.Images and ebi attached.X-rays unavailable.Products not returning: hospital keeps.320-36-00 36mm humeral liner +0 unconstrained (b)(6).Udi number (b)(4).510(k) number k093275.Product code kws, kwt.Concomitants: (b)(6) - gps implant kit v2 (b)(6).320-35-02 - small superior augment glenoid plate (b)(6).320-20-26 - eq rev compress screw lck cap kit, 4.5 x 26mm (b)(6).320-20-26 - eq rev compress screw lck cap kit, 4.5 x 26mm (b)(6).300-01-11 - equinoxe, humeral stem primary, press fit 11mm (b)(6).320-20-00 - eq reverse torque defining screw kit (b)(6).531-55-88 - ergo gps 3.2mm drill kit sterile (b)(6).320-20-18 - eq rev compress screw lck cap kit, 4.5 x 18mm (b)(6).531-78-20 - shouldr gps hex pins kit (b)(6).320-15-05 - eq rev locking screw (b)(6).320-31-36 - glenosphere, 36mm (b)(6).320-10-00 - equinoxe reverse tray adapter plate tray +0 (b)(6).320-20-18 - eq rev compress screw lck cap kit, 4.5 x 18mm (b)(6).
 
Additional Manufacturer Narrative
The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Additional Manufacturer Narrative
D11 concomitant devices: a10012 - gps implant kit v2 09003122202, 320-35-02 - small superior augment glenoid plate a006673, 320-20-26 - eq rev compress screw lck cap kit, 4.5 x 26mm a155072, 320-20-26 - eq rev compress screw lck cap kit, 4.5 x 26mm a166808, 300-01-11 - equinoxe, humeral stem primary, press fit 11mm a193925, 320-20-00 - eq reverse torque defining screw kit a266892, 531-55-88 - ergo gps 3.2mm drill kit sterile a339914, 320-20-18 - eq rev compress screw lck cap kit, 4.5 x 18mm a343046, 531-78-20 - shouldr gps hex pins kit a365632, 320-15-05 - eq rev locking screw a369536, 320-31-36 - glenosphere, 36mm a373322, 320-10-00 - equinoxe reverse tray adapter plate tray +0 a405792, 320-20-18 - eq rev compress screw lck cap kit, 4.5 x 18mm s382797.
 
Event or Problem Description
It was reported that this patient's right knee was revised.Since the index surgery, the patient had a colostomy placed.At some point after, patient developed an infection.It is unclear if her infection is related to the colostomy procedure.Patient had an abscess within the soft tissue of her deltoid.Surgeon drained the abscess and performed a one-stage revision.He explanted all the original hardware and implanted a new small superior augment baseplate, blue, red, and black screws and caps, 40 expanded sphere, locking screw, cemented 12mm stem, 0 tray, reverse torque screw and a 2.5mm liner.Original liner had some wear already.Patient was last known to be in stable condition following the event.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: b, c, d, e.The reason for the revision reported was likely the result of infection as reported.The cause of the reported infection cannot be conclusively determined; however, it may be related to the patient¿s previous colostomy procedure and/or underlying patient condition.The prosthesis wear noted on the explanted humeral liner is likely the result of impingement between the humeral liner and native glenoid bone.However, this cannot be confirmed as devices were not returned for evaluation and no radiographs were provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
145-DEG PE 36MM HUM LINER +0
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key20893579
Report Number1038671-2024-04736
Device Sequence Number15382548
Product Code KWS
UDI-Device Identifier10885862086617
UDI-Public10885862086617
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K093275
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number320-36-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2024
Supplement Date Received by Manufacturer12/10/2024
09/26/2025
Initial Report FDA Received Date12/10/2024
Supplement Report FDA Received Date12/10/2024
09/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Patient SexFemale
-
-