| Model Number |
OTV-S7H-1D-F08E |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: (b)(6) hospital.E2/e3: no information available for the occupation of the initial reporter or whether they are a healthcare professional.The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that during set up, the subject device had poor image quality.There was no patient involvement.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Updated fields: d8, h2, h3, h4, h6, h11 the device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunction was identified during the device evaluation: charged coupled device failure, cable waterproofing defects and cable scratches.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: due to the charged coupled device failure, there is image defect.The most probable cause of the complaint is traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|