AIZU OLYMPUS CO., LTD. EVIS EXERA III BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-P190
Medical Device Problem Code Microbial Contamination of Device (2303)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported that the scope tested positive for an unexpected contamination.The issue was found during a routine culture of the scope.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Event or Problem Description
It was reported that, during reprocessing, the scope tested positive for 40 colony forming units (cfus) of unspecified microbes.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the results of third-party testing, the device evaluation and the final investigation.Olympus provided the following result of the culture test, performed at the third-party labs: sampling date: (b)(6) 2024.Sampling from: all channels.Cfu: <1cfu.Bacterial identification: n/a.The device was evaluated where no abnormalities were found that could have led to the reported event.Despite good faith attempts the user cleaning disinfection and sterilization (cds) processes were not shared.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the ifu before repair, the results conformed to the regulation's recommendation.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III BRONCHOVIDEOSCOPE
Common Device Name
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20894551
Report Number9610595-2024-40456
Device Sequence Number7843121
Product Code EOQ
UDI-Device Identifier04953170434983
UDI-Public04953170434983
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K201758
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBF-P190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/13/2024
Supplement Date Received by Manufacturer07/30/2025
Initial Report FDA Received Date12/10/2024
Supplement Report FDA Received Date08/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/18/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-