Model Number 978B128
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/20/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3 analysis of the returned lead revealed the distal end of the lead was stretched.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was implanted with an implantable neurostimulator (ins) for urinary dysfunction (incontinence, urgency, and/or frequency).It was reported that the lead was damaged.The inner wire departed from the sheath during surgery.Cause was not determined.No troubleshooting was per formed.The issue was confirmed resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key20894592
Report Number2649622-2024-33305
Device Sequence Number18071457
Product Code EZW
UDI-Device Identifier00763000596910
UDI-Public00763000596910
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P970004
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/24/2026
Device Model Number978B128
Device Catalogue Number978B128
Device Lot NumberVA319XE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2024
Initial Report FDA Received Date12/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age76 YR
Patient SexFemale
-
-