MEDOS INTERNATIONAL SARL HD C-MNT SCP,2.7,70,75,MITEK; ARTHROSCOPE Back to Search Results
Catalog Number 242277
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during functional testing, the hd c-mnt scp,2.7,70,75,mitek device had a blurry image during functional test.The adjustable zoom coupler feature was non functional.During in-house engineering evaluation, it was determined that the device was broken in to two pieces/shaved/delaminated.There were no adverse patient consequences.There were no delays to the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: outer tube damaged, distal tip distal tip damaged dings/scratched, illumination distal tip fiber damaged, particulate, optics particulate under distal lens particulate in system, optical system, optical components prism loose proximal cover glass damaged residue broken (2+ pieces) : chipped/shaved/delaminated, visual : deformed/bent, visual : foreign substance/debris/cleaning/sterilization.Per service reports, this complaint was confirmed.The defective parts need to be replaced to resolve the issues.Based on the investigation findings, the faulty parts were identified as the root cause for the device failure during the service evaluation.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: per service report, this complaint was confirmed.The optical, tube, lenses were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD C-MNT SCP,2.7,70,75,MITEK
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle 02400
SZ   02400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key20894973
Report Number1221934-2024-04399
Device Sequence Number8710353
Product Code HRX
UDI-Device Identifier10886705031917
UDI-Public(01)10886705031917
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K080560
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number242277
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2024
Supplement Date Received by Manufacturer01/29/2025
Initial Report FDA Received Date12/10/2024
Supplement Report FDA Received Date01/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-