The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.Based on the information received, the investigation determined that the reported balloon rupture appears to be related to operational context.There was no leak noted during preparation, which suggests a product quality issue did not contribute to the reported difficulty.In this case, it is likely that the balloon interacted with the heavily calcified anatomy such that the balloon became damaged and subsequently ruptured during inflation.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
|
It was reported that the procedure was to treat a 90% stenosed lesion in the superficial femoral artery (sfa) with heavy calcification and heavy tortuosity.The 5x40mm armada 35 balloon dilatation catheter (bdc) was advanced to the target lesion and the balloon was inflated; however, the balloon ruptured at 2 atmospheres (atm) during the first inflation.Therefore, the bdc was removed from the patient.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
|