• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL CEREPAK HELIFORM; CEREPAK HELIFORM XL DETACHABLE COIL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL CEREPAK HELIFORM; CEREPAK HELIFORM XL DETACHABLE COIL SYSTEM Back to Search Results
Catalog Number SHD180415
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/05/2024
Type of Reportable Event Malfunction
Event or Problem Description
As reported by the field, during a coil embolization, in a very large basiliar aneurysm, the physician was unable to detach a heliform xl 4mm x 15 cm coil (shd180415, 31098462) and a uniform 4mm x 11.5 cm coil (fcx100412, 31207070).Each coil was attempted initially with a mechanical detachment handle (mdh1, lot unknown).After that was unsuccessful, the manual break was attempted without success.Each coil was taken out of the aneurysm without issues and another coil was used with the same microcatheter (plato 17 straight).With each coil, the microcatheter outside the body was straight.No negative impact on patient and no significant delay in treatment.
 
Additional Manufacturer Narrative
Product complaint(b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section d2b ¿ procode: krd/hcg.The device was discarded; therefore, no further investigation can be performed.A manufacturing record evaluation was performed for the finished device 31207070 number, and no non-conformances related to the malfunction were identified.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the device history record review, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be determined; however, there are circumstances of the procedure that may have contributed to the reported failure.The instructions for use (ifu) contain the following recommendations: 1.Remove the detacher from the protective packaging and place it within the sterile field.The detacher is packaged separately for single patient use only.2.Confirm again under fluoroscopy that the coil alignment marker of the delivery tube creates a ¿t¿ with the proximal marker of the microcatheter.3.Verify that the rhv is firmly locked around the delivery tube to ensure that the coil does not move during connection process.Ensure the delivery tube is straight between the rhv and the detacher.4.Hold the delivery tube approximately 3 inches from proximal end and bring the detacher over the delivery tube until a firm stop is felt.5.Maintain the delivery tube position within the detacher and use a thumb to pull back the slider on the detacher.6.Continue to pull on the slider until end of travel or a click is heard.Gently release the slider until it returns to starting position or a second click is heard.Remove detacher by slowly withdrawing it from the delivery tube.Note: detacher must remain in sterile field until the procedure is completed.7.Fluoroscopically verify that the coil is detached by gently pulling back the delivery tube approximately 1cm.If the coil has detached, remove the delivery tube from the microcatheter.8.If the coil has not detached, repeat steps 3-7.If 2 unsuccessful detachments occur, discard the detacher and proceed to the manual break section.9.Repeat steps 1-8 until the procedure is completed.Discard the detacher.Caution: always perform fluoroscopic verification of detachment prior to withdrawing the delivery system fully.Failure to do so may lead to an embolic complication.10.After detaching the coil, remove the delivery tube from the microcatheter and discard.11.Repeat the above sequence for all additional coils until the procedure is complete.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The company is seeking this information through the event investigation.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CEREPAK HELIFORM
Common Device Name
CEREPAK HELIFORM XL DETACHABLE COIL SYSTEM
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS, INC.
calle circuito int norte 1820
parque industrial salvacar
juarez 32574
MX   32574
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key20895311
Report Number3008114965-2024-01201
Device Sequence Number20813549
Product Code KRD
UDI-Device Identifier10886704083474
UDI-Public(01)10886704083474(17)260531(10)31098462
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K220040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Catalogue NumberSHD180415
Device Lot Number31098462
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2024
Initial Report FDA Received Date12/10/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/21/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MECHANICAL DETACHMENT HANDLE; PLATO® 17 MICROCATHETER (SCIENTIA VASCULAR)
Patient SexUnknown
-
-