• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number 620-0221011-01
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device would not power up.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Additional Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation.The customer's report was verified and attributed to the monitor board.The monitor board was replaced to remedy the report.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key20896207
Report Number1220908-2024-04373
Device Sequence Number9043737
Product Code MKJ
UDI-Device Identifier00847946018726
UDI-Public(01)00847946018726
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P160022
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number620-0221011-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2024
Initial Report FDA Received Date12/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/01/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-