| Catalog Number |
6260-9-240 |
| Medical Device Problem Code |
Corroded (1131)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.
|
| |
|
Event or Problem Description
|
|
It was reported by counsel that claimant underwent left tha on (b)(6) 2009 and was implanted with an lfit anatomic v40 femoral head and accolade tmzf hip stem.He was revised on (b)(6) 2023 allegedly due to corrosion.
|
| |
|
Search Alerts/Recalls
|