• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. MCRYL+ UD 18IN 3-0 S/A PS-2 PRM MP; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. MCRYL+ UD 18IN 3-0 S/A PS-2 PRM MP; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number MCP497G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 device not returned.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The following information was requested but unavailable: ¿ were there any patient consequences? ¿ did any needle piece fall into the patient? if yes, ¿ was the needle piece(s) retrieved during the same procedure? ¿ were x-rays taken to locate the needle piece(s)? ¿ what measures were taken to retrieve the broken piece? ¿ was there any additional tissue damage as a result of searching for the needle piece? ¿ if not retrieved, in what tissue structure the broken needle was retained? is there any plan in place to remove the needle piece in the future?¿ please perform and document the follow-up attempt for product return.
 
Event or Problem Description
It was reported that a patient underwent a vaginal delivery on (b)(6) 2024 and suture was used.During the procedure, the needle broke when sewing.Changed to another one to complete the surgery.There were no adverse patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCRYL+ UD 18IN 3-0 S/A PS-2 PRM MP
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key20897009
Report Number2210968-2024-13169
Device Sequence Number8942926
Product Code GAM
UDI-Device Identifier10705031046603
UDI-Public(01)10705031046603(17)250930(10)TLMBCR
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K050845
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Catalogue NumberMCP497G
Device Lot NumberTLMBCR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/13/2024
Supplement Date Received by Manufacturer12/18/2024
Initial Report FDA Received Date12/10/2024
Supplement Report FDA Received Date01/14/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-