| Model Number |
99351 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H6: 3038: catheter.The product involved in the report has not been returned.A review of the device history record is in-progress.All information reasonably known as of 10 dec 2024 has been included in the health authority report.Should additional information be obtained a follow-up health authority report will be provided.The information provided by airlife represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and is identified as complaint - (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a airlife product is defective or causes serious injury.
|
| |
|
Event or Problem Description
|
|
It was reported, that after tracheal suction, air entered into the sleeve and it inflated.The catheter became difficult to insert.
|
| |
|
Event or Problem Description
|
|
It was reported, that after tracheal suction, air entered into the sleeve and it inflated.The catheter became difficult to insert.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: 3038: catheter.The product involved in the report has not been returned.A review of the device history record is in-progress.All information reasonably known as of 10 dec 2024 has been included in the health authority report.Should additional information be obtained a follow-up health authority report will be provided.The information provided by airlife represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and is identified as complaint: (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a airlife product is defective or causes serious injury.The complaint of " after tracheal suction, air entered into the sleeve and it inflated.The catheter became difficult to insert." regarding part: 2160-4j was not confirmed per sample evaluation, therefore, the root cause could not be determined.A risk assessment was performed and the ultimate risk was determined to be low which does not require the initiation of a capa.There has been one other not confirmed complaint regarding the same part and a similar issue within the 24 months preceding this reported event.A resolution letter was sent to the customer.Complaints will continue to be monitored for possible trends.
|
| |
|