MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3501690 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Breast Discomfort/Pain (4504)
|
| Date of Event |
11/21/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: n/a.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation revision procedure with a mentor smooth round moderate profile 575cc saline breast prosthesis that deflated post implantation.Deflation of the patient¿s left breast prosthesis was diagnosed during a phone consultation.Additionally, the patient developed pain in her left breast and shoulder.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.
|
| |
|
Additional Manufacturer Narrative
|
|
On 06-feb-2025, the mentor failure analysis lab received the device for evaluation.Additionally, mentor became aware that the patient underwent explantation on (b)(6) 2025.On 07-feb-2025, mentor received additional information about the event.The patient had both breast prostheses explanted and they were replaced with mentor memorygel boost breast implant 630cc gel breast prostheses.On (b)(6) 2025, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, the patient experienced deflation in the breast implant.In addition, the patient presented with breast pain.The product was returned to mentor for evaluation.Mentor conducted visual inspection and microscopic examination of the returned device.Visual analysis of the returned device revealed that the breast implant had a crease/fold on the posterior view.In addition, a tear was found within the crease/fold, measuring approximately 0.5 cm.A microscopic examination was performed and no instrument damage was observed at the edges of the rupture.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in a deflation in a later time.Breast implants may also simply wear out over time.Most women undergoing augmentation or reconstruction with a mammary prosthesis will experience some pain postoperatively.While pain normally subsides in most women as they heal from surgery, it can become a chronic problem in other women.Chronic pain can be associated with a variety of factors.Surgeons should instruct their patients to inform them if there is significant pain or if pain persists.Pain is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Reason for device explant and/or reoperation: breast prosthesis deflation, breast pain.Manufacturer¿s reference number: (b)(4).
|
| |
|