The healthcare professional reported that during an endovascular coil embolization procedure to treat a subarachnoid hemorrhage (sah), after the excelsior® sl-10® microcatheter (stryker) was placed into the target aneurysm, the 2.00mm x 4.00cm galaxy g3 mini (glm920040 / 30830566) was used as the first coil.The coil was rewound inside the aneurysm in an attempt to retrieve it, but it could not be resheathed as the introducer sheath got damaged.A continuous flush was maintained through the microcatheter.There was no report of any negative patient impact.On 10-dec-2024, additional information was received.Per the information, the reason that led to the coil needing to be resheathed was ¿positioning difficulty.The frame was unable to be formed properly, so [the coil was rewound several times].¿ during the coil rewinding, the microcatheter came out of the aneurysm and the coil had to be resheathed.The introducer was split.There was no resistance at any during the advancement of the coil through the microcatheter.The information stated no neck remodeling was used.The microcatheter was re-introduced into the patient.Based on the additional information received on 10-dec-2024, the event has been deemed usfda reportable under 21 cfr 803 with a classification of ¿malfunction.¿.
|
|
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, sex, gender, weight, race, and ethnicity were not provided.Section d.2b: procode is krd/hcg.Section e.1: the initial reporter phone is not available / reported.Based on complaint information, the device is not available to be returned for analysis.A review of manufacturing documentation associated with this lot (30830566) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no internal actions related to device manufacture or inspection.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.With the limited information available and without the product available for analysis, the reported issue documented in the complaint could not be confirmed.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be conclusively determined; however, it is possible that circumstances of the procedure and / or device manipulation / interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered later.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by johnson & johnson medtech, or its employees that the report constitutes an admission that the product, johnson & johnson medtech, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|