Model Number MR8-AA07
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis: evaluation confirmed overheating.The device was tested and the maximum temperature was measured to be >109°f.The likely cause of failure could not be determined.It was also noted noise, laser markings faded, housing nut is completely stuck and was heavily damaged.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Repair request initiated for device with the report of overheating.No patient impact reported.Repair request escalated to product event due to customer indication that this report is a complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATT MR8 ANGLED
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key20900203
Report Number1625507-2024-01184
Device Sequence Number15689128
Product Code HBB
UDI-Device Identifier00643169912380
UDI-Public00643169912380
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K183515
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR8-AA07
Device Catalogue NumberMR8-AA07
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2024
Initial Report FDA Received Date12/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/13/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1