• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, FOOTSWITCH; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, FOOTSWITCH; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065751762
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Intraocular Pressure Increased (1937); Posterior Capsule Opacification (4807)
Date of Event 11/14/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Event or Problem Description
A physician reported that during cataract surgery, an ophthalmic system footswitch displayed message and pedal failure occurred after incision.The patient experienced posterior capsular opacification and intraocular hypertonia, which required additional intervention the surgery was not completed and the patient transferred to another hospital.The current patient status was unknown.This report pertains to ophthalmic footswitch.
 
Event or Problem Description
Additional information received indicates that the patient's condition is improving.There was no inflammation.
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Additional Manufacturer Narrative
Additional information was provided in sections h.6 and h.11.Service history was reviewed for the system.A service record relevant to the reported complaint was found.Service record (sr) was opened on to address the reported event.The sr was closed, and no work was performed.There has been no service requested by the customer in response to the reported event.Therefore, the customer reported event cannot be confirmed.Based on the information obtained, the root cause of the reported events is inconclusive.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.Based on the information obtained, the root cause of the reported events is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, FOOTSWITCH
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20901257
Report Number2028159-2024-01925
Device Sequence Number20084620
Product Code HQC
UDI-Device Identifier00380657517626
UDI-Public(01)00380657517626(17)270620(11)170505(10)NA
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K121555
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number8065751762
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/14/2024
Supplement Date Received by Manufacturer12/09/2024
01/28/2025
Initial Report FDA Received Date12/11/2024
Supplement Report FDA Received Date01/06/2025
02/11/2025
Date Device Manufactured05/05/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CENTURION VISION SYSTEM
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age41 YR
Patient SexFemale
-
-