| Catalog Number |
8065751762 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); Posterior Capsule Opacification (4807)
|
| Date of Event |
11/14/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Event or Problem Description
|
|
A physician reported that during cataract surgery, an ophthalmic system footswitch displayed message and pedal failure occurred after incision.The patient experienced posterior capsular opacification and intraocular hypertonia, which required additional intervention the surgery was not completed and the patient transferred to another hospital.The current patient status was unknown.This report pertains to ophthalmic footswitch.
|
| |
|
Event or Problem Description
|
|
Additional information received indicates that the patient's condition is improving.There was no inflammation.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in sections h.6 and h.11.Service history was reviewed for the system.A service record relevant to the reported complaint was found.Service record (sr) was opened on to address the reported event.The sr was closed, and no work was performed.There has been no service requested by the customer in response to the reported event.Therefore, the customer reported event cannot be confirmed.Based on the information obtained, the root cause of the reported events is inconclusive.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.Based on the information obtained, the root cause of the reported events is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|