EXACTECH, INC. TRULIANT TIB IMP CR INS STD SZ 2, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-022-47-2009 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Osteolysis (2377); Metal Related Pathology (4530); Implant Pain (4561)
|
| Date of Event |
09/05/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Report numbers: 1038671-2024-03688, 1038671-2024-03689 d10 concomitants: (b)(6) 02-022-47-2009 - truliant tib imp cr ins std sz 2, 9mm.(b)(6) - 00-5071-129-00 - 105x19.5x1.27mm.E1: phone number (b)(6).Multiple mdr reports were filed for this event, please see associated reports: 1038671-2024-03688, 1038671-2024-03689.The reason for the patient¿s second revision reported was likely the result of loosening between the tibial and femoral implants and the bone after almost 11 years of implantation.Osteolysis, and possibly loosening, on both the tibial and femoral sides appears to have been present at the time of the first revision.As the tibial insert had been recently revised, the articular surface damage is likely consistent with third-body wear secondary to loosening and non-articular damage secondary to femoral subsidence.However, this cannot be confirmed as the devices were not returned.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
As reported, approximately 10 months post the previous revision surgery, the patient had knee pain and the surgeon decided to revise again because the femoral and tibial components had been found to be loose.The patient experienced implant pain and osteolysis.Metallosis was found and all proliferative tissue removed, then all components were replaced to competitor devices.There was no reported breakage of a device or surgical delay/prolongation.The patient was last known to be in stable condition following the event.No further issues or complications were reported.No additional information is available.
|
| |
|
Search Alerts/Recalls
|
|
|