| Model Number |
ANGLED REAMER SLEEVE, 10° |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024, it was reported by a sales representative via (b)(4) that an ar-9597-10 reamer sleeve and ar-9676 drive shaft are damaged non-specifically.This occurred during use in a case with no patient effect.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Complaint allegation is confirmed.One unpackaged ar-9597-10 serial/batch number (b)(6) was received for investigation.Functional testing was performed with a known good mating part, ar-9676 angled reamer and found the devices would not twist together and some resistance was felt.Visual inspection noted signs of wear on the device: striation marks and nicks along the shaft.As well as damage on the proximal and distal end.The most likely cause for the reported failure can be attributed to misuse due to interference with the mating instrument.
|
| |
|